Mediana P10 Manuale di Servizio Pagina 23

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 22
17
b. Press the Print Button when all the parameter signals display normally.
c. Verify the parameter values and waveforms are printed out with 25 mm/s.
d. Set Print Speed to 50 mm/s.
e. Press the Print Button when all the parameter signals display normally.
f. Verify that the parameter values and waveforms are printed out with 50 mm/s.
g. Verify the two ECG waveforms are printed out.
h. Repeat this test for other selections.
8. Test #5: Print-On-Alarm
a. Set Print-On-Alarm to ON via the Alarm Limits Menu.
b. Set the Heart Rate of your ECG simulator to 30bpm.
c. Verify Low Heart Rate/Pulse Rate limits violated” alarm is activated and
the parameter values and waveforms are printed out.
Note: If no printer is installed in the monitor, Print Mode, Print Speed and Print On Alarm
will not display.
Note: If there is no printer paper left or printer paper feeds improperly, the monitor will
display an alarm message.
Vedere la pagina 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 ... 109 110

Commenti su questo manuale

Nessun commento